logo

VICI DBS S.r.l.
VICI AG International
VICI DBS USA

H2-Sensor

The VICI DBS H2 - Sensor provides a real-time bar graph display of the concentration of hydrogen in the GC oven. If there is leakage to the threshold level, an audible alarm sounds and the display flashes.  If the leak continues until the upper limit of hydrogen concentration is reached, the system will stop the VICI DBS hydrogen generator. An optional feature is a valve which will activate an inert gas purge of the GC oven when the upper limit of hydrogen is reached.  When normal levels are restored, operations can be restarted.

The following option is available:
• Purge gas valve

Detection range

0 – 25 % LEL (0 – 1.0 % by vol. H2)

0.4 % by vol.

H2 Optical alarm: shows you the percentage of LEL, acoustic alarm: starts with intermittent beeps

1.0 % by vol.

H2 Optical alarm: shows you the percentage of LEL, acoustic alarm continues

Size of Sensor Box

120 mm (w) x 172 mm (d) x 70 mm (h)

Electrical requirements

100-240 V, 50 / 60 Hz, 1.0 A

Optional

ON / OFF valve for inert gas or for blocking the H2 line

Acoustical alarm

YES

Visual alarm

YES