Mistral Evolution Family

The Mistral-Evolution nitrogen gas generator is specifically designed and built to meet the flow, purity, and pressure requirements of today’s LCMS systems.  It can also be used for the evaporation of solvents in samples being analyzed. The simple but well-proven pressure swing adsorption (PSA) technology is used to separate nitrogen from the other gases present in compressed air.

ADVANTAGES
The Mistral-Evolution family is optimized to generate nitrogen of constant purity for improved performance of analytical instruments, resulting in an overall  improvement in lab productivity. It is especially suited for LCMS, which requires a continuous supply of pure nitrogen which usually cannot be provided by gas tanks. With no gas tanks, there are no expenses related to gas supply, transportation, storage, and handling.  As an added benefit, the pure nitrogen produced at low pressure and room temperature eliminates the risks associated with high pressure gas bottles and the handling of liquid nitrogen.

The Mistral-Evolution is easy to install.  A 220V-50Hz (optional 110V-60Hz) power source is all that is required. The low noise of the instrument allows installation near the LCMS, but it can be located up to 30 meters away.

MODELS AVAILABLE
• Mistral-Evolution 10 Nl / min
• Mistral-Evolution 15 Nl / min
• Mistral-Evolution 25 Nl / min
• Mistral-Evolution 35 Nl / min
• Mistral-Evolution 40 Nl / min
• Mistral-Evolution Hybrid for the LCMS AB SCIEX
• Mistral-Evolution Gas-Station x Agilent AES 4100MP

MISTRAL EVOLUTION
The Mistral LCMS Evolution (PSA technology), which includes an integral oil-free air compressor, delivers a continuous stream of pure nitrogen gas of 98.5% at a flow rate of 40 Nl/min. It uses the pressure swing adsorption (PSA) system that removes oxygen, carbon dioxide, and water from compressed air. The resulting stream of pure nitrogen is ideal for LCMS techniques and other inert gas applications.

MISTRAL EVO HYBRID X LCMS AB SCIEX
The Mistral EVO Hybrid nitrogen generator was specifically designed to meet the flow, purity, and pressure requirements of the LCMS of AB SCIEX. The simple and proven pressure swing adsorption (PSA) technology is employed to remove nitrogen from compressed air. Selective molsieves (CMS) adsorb remaining oxygen and traces of humidity.

MISTRAL EVO GAS-STATION X AGILENT AES 4100MP
The new Mistral EVO Gas-Station is the global dedicated solution for the Microwave Agilent 4100MP system. The well-known and proven pressure swing adsorption (PSA) technology is used to simultaneously generate nitrogen and dry air -- the nitrogen with a purity >99.5% at up to 25 Nl/min, and the air at up to 40 Nl / min, suitable for POP and EGM. This combination allows the system to stabilize quickly and be ready faster. Thanks to a specialized insulation, the Mistral EVO Gas-Station can be installed in the laboratory near the instrument or underneath the laboratory bench.

The following option is available:
• 10 liter tank

Mistral Evolution

Model

MISTRAL EVO 10L

MISTRAL EVO 25L

MISTRAL EVO 35L

MISTRAL EVO 40L

Flow

10 Nl/min

25 Nl/min

35 Nl/min

40 Nl/min

Purity N2

>99.5 %

>99.5 %

>99 %

>99 %

Pressure Nitrogen Outlet

max 8 bar (116 psi)

Voltage

110V / 60Hz - 230V / 50Hz

Weight

90 Kg net

Consumption

1800 W

Pressure accuracy

0.1 bar (± 0.5 %)

Display

Graphic Display, 128 x 64 px

Protection

IP2x

Temperature

From 15 °C to 40 °C

Noise

< 60 dB(A)

Alarm relay

250 VAC, 16 A max

Outlet Nitrogen

Outlet, rapid fitting for 6mm OD tube

Dimensions (mm)

482 x 835 x 641 (L x P x A)

Certification

CE, CSA, FCC

 

Mistral Evolution Hybrid

N2 Flow

Up to 12 Nl / min at 80 psi

Source Air Gas

Up to 25 Nl / min at 100psi

Exhaust Air Gas

Up to 10 Nl / min at 60 psi

Purity

> 99%

Voltage

110V 60Hz - 220V 50Hz

Weight (kg)

90

Consumption

1800W

Pressure Accuracy

0.1 bar (± 0.5%)

Display

Graphic Display, 128x64px

Protection

IP2x

Temperature

15 °C – 40 °C

Air dewpoint at operating pressure

-40 °C

Background noise

< 60dB

Alarm relay

250VAC, 16Amp max

Connections

6mm OD tube rapid fittings

Dimensions (mm)

482 x 835 x 641 (L x P x A)

Certification

CE, CSA, FCC

 

Mistral Evolution Gas-Station

N2 Gas Flow

Up to 25 Nl/min at 8 bar max

Air Gas Flow U

p to 40 Nl/min at 8 bar max

Purity

> 99,5%

Voltage

110V 60Hz - 220V 50Hz

Weight (kg)

110 net

Consumption

2500W max

Pressure accuracy

0.1 bar (± 0.5%)

Display Graphic Display,

128x64px

Protection

IP2x

Temperature

15 °C – 40 °C

Air dewpoint at operating pressure

-40 °C

Background noise

< 60dB

Alarm relay

250VAC, 16Amp max

Connections

6mm OD tube rapid fittings

Dimension (mm)

482 x 1235 x 641 (L x P x A)

Certification

CE, CSA, FCC