N2 Tower Plus

The new N2-Tower Plus series of nitrogen generator produce high purity nitrogen for all major applications in analytical laboratories. Nitrogen generators eliminate the problems associated with the handling of high pressure gas tanks and reduce the cost of high purity nitrogen immediately.

The N2-TOWER Plus series of nitrogen generators use air from a separate compressor and remove the oxygen to less than 10ppm (hydrocarbons to <0.1ppm is optional). The guaranteed flow rates of pure nitrogen are 500 ml/min, 750 ml/min, 1300 ml/min or 4000 ml/min depending on the generator module.

SPECIFICATIONS
• LED-display for status
• Maintenance control light
• Alarms for catalyst duration, catalyst temperature, and pressure
• Simple filter removal without the need to open the instrument

APPLICATIONS
N2-TOWER Plus series nitrogen generators are ideal for the method development and routine analysis by GC, GC-HS, ICP, HPLC-ELSD.
• GC carrier gas and make-up gas
• ICP / HPLC-ELSD
• GC / FTIR
• Incubators
• TGA-DSC

BENEFITS

The constant high purity of the nitrogen provides stable conditions for analytical systems and guarantees highly reproducible results. The purchase price for the nitrogen generator is generally recovered in about one year, since there is no longer the extra cost involved in handling, transporting, storing, and purchasing high pressure gas tanks. The installation is very simple and quick, as is the routine maintenance.

OPERATING PRINCIPLE
Compressed air from an outside source (not included) passes through a charcoal filter for pre-cleaning before the oxygen is removed by the double column pressure swing adsorption (PSA) system. One of the PSA columns is active while the second one is regenerated. The removed gases, like oxygen, are released back to the atmosphere.

With an optional catalyst, other trace impurities such as hydrocarbons and CO are removed to levels <0.1 ppm. An additional filter is installed to remove particulates down to 0.5 microns.  For the complete removal of all hydrocarbons, an optional catalytic oven is available. This will guarantee trouble-free operation of FID detectors in GC.

The following option is available:
• Catalytic oven

Models

N2-TOWER

500

N2-TOWER

750

N2-TOWER

1300

N2-TOWER

4000

Flow Nml / min.

500

750

1300

4000

Purity

99.999%

99.995%

99.99%

99%

Delivery pressure

Up to 5 barg (75 Psig)

Input pressure for model without compressor

7 min - 10 barg max (110-145 PSI) / Non condensing and oil free Air

N2 purity

0.1 ppm Total Hydrocarbons (if Option is installed) < -50 °C DP (-58 °F)

Sound level

Noiseless (external compressor)

Particle filtration level

50 μm

Fittings

Inlet: 1 / 8’’ G thread with 6mm tube rapid fitting

Outlet: 1/8” G thread with 4mm tube rapid fitting

User interface

Pressure setting, system status, LED RGB display

Dimensions (mm)

140W x 630D x 490H

Power consumptions

20W - 270W with catalytic oven

Power requirements

110-120V 60Hz / 220-240V 50Hz